3-10
Section 3-4 - Completing the Installation
3-4-4
Power Off/ Shutdown
NOTE:
After turning off a system, wait at least ten seconds before turning it on again. The system may not be
able to boot if power is recycled too quickly.
3-4-4-1
Back-end Processor Power Down
1.) Press the
ON/OFF
Stand-By switch left below the Control Panel once.
Figure 3-9
3-4-4-2
Scanner Shutdown
1.) Press the
ON/OFF
Stand-By switch left below the Control Panel once.
2.) Switch OFF the Circuit Breaker at the rear of the system.
NOTE:
The mains outlet of the system for peripheral auxiliary equipment are commonly switched with
the Standby switch. So the auxiliary equipment need not to be switched ON/OFF separately.
3.) Disconnect the Mains Power Cable if necessary.
For example:
Repairing the system.
4.) Press once on the brakes to block the front wheels (brakes on front wheels under the foot rest).
5.) Prior to disconnect a probe freeze the image.
6.) Open the right-hand side door, remove the probe cable from the cable holder and close the door.
7.) Move the probe locking handle counterclockwise. Pull the connector straight out of the probe port.
Figure 3-10 Circuit Breaker and Power Cable on Back of Voluson® 730
CAUTION
If a probe is disconnected while running (Write-Mode) a software error may occur.
In this case switch the unit OFF (perform a reset).
Circuit Breaker
Mains Power Cable
Содержание Voluson 730
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL iii ...
Страница 8: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL vi ...
Страница 24: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL xxii ...
Страница 268: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL IV ...
Страница 269: ......