Chapter 3 Installation
3-7
3.) Disconnect the GND ground-cable from the back of the rear cover plate.
4.) Connect all Peripherals.
For connection schemes of Peripherals refer to Chapter 20 in the Basic User Manual of Voluson® 730.
5.) Connect the GND ground-cable at the rear cover plate.
6.) Mount rear cover plate.
-
Reattach screw and washers.
-
Plug-in caps.
3-3-2
Verify Customer Order
Compare items received by the customer to that which is listed on the delivery order. Report any items
that are missing, back ordered or damaged.
3-3-3
Physical Inspection
3-3-3-1
System Voltage Settings
Verify that the scanner is set to the correct voltage.
The Voltage settings for the Voluson® 730 Scanner is found on a metal plate, on the rear of the system.
3-3-4
EMI Protection
This unit has been designed to minimize the effects of Electo-Magnetic Interference (EMI). Many of the
covers, shields, and screws are provided primarily to protect the system from image artifacts caused by
this interference. For this reason, it is imperative that all covers and hardware are installed and secured
before the unit is put into operation.
Figure 3-6 GND ground-cable
Figure 3-7 Identification Plate Example
WARNING
WARNING
CONNECTING A Voluson® 730 SCANNER TO THE WRONG VOLTAGE LEVEL WILL
MOST LIKELY DESTROY THE SCANNER.
GND ground cable
Содержание Voluson 730
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL iii ...
Страница 8: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL vi ...
Страница 24: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL xxii ...
Страница 268: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL IV ...
Страница 269: ......