Chapter 7 Diagnostics/Troubleshooting
7-23
Execute privileged instruction
restart the system
FallSmooth Out Of Range
Press ok and save this user-setting once again
File Could not CRC Check
Load volume files from other storage medium.
File CRC Error
Load volume files from other storage medium.
File CRC Missing
Load volume files from other storage medium.
File Data Missing
Load volume files from other storage medium.
File Datalength Not Consistent
Load volume files from other storage medium.
File Decompress Error
Load volume files from other storage medium.
File Decompress method Unknown
Load volume files from other storage medium.
File End Error
Load volume files from other storage medium.
File Execute Error
Load volume files from other storage medium.
File Memory Missing
Load volume files from other storage medium.
File Not Found
Load volume files from other storage medium.
File Not Found
Load another Backup from CD,…. (restart or reinstall the
Application)
File Not Found
restart, install latest software version
File Pos
Load volume files from other storage medium.
File Read Error
Load volume files from other storage medium.
File Type Unknown
Load volume files from other storage medium.
File Volume size not consistent
Load volume files from other storage medium.
FLT: Denormal operand
restart the system
FLT: Divide by zero
restart the system
FLT: Invalid operation
restart the system
FLT: Stack overflow
restart the system
FLT: Underflow
restart the system
Hardware doesn't support CW-mode
pencil probe + CW-Hardware not available->HW problem
hardware error on
restart the system
HardwareRelatedSoftware_Windows in write have different
ProbeAcousticUnitIDs
disconnect all connected probes and connect they anew, if
not ok restart the system
HardwareRelatedSoftware_Windows in write have different
ProbeScanFuncIDs
disconnect all connected probes and connect they anew, if
not ok restart the system
IBegrenzer.cpp Bshots TxMultiFocus problem
restart the system
IBegrenzer.cpp Mshots TxMultiFocus problem
restart the system
Error Messages
Actions
Содержание Voluson 730
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL iii ...
Страница 8: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL vi ...
Страница 24: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL xxii ...
Страница 268: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL IV ...
Страница 269: ......