Chapter 3 Installation
3-15
3-5-2-1
Install the UP-D21MD printer software/driver
(cont’d)
6.) Use the BROWSE button to search the following path
C:\Utilities\PrinterDriver\SonyUP-D21MD
see:
Figure 3-13
and then click OK.
7.) Confirm the correct path and click NEXT to install the driver. All necessary files are copied.
8.) The Message:
Digital Signature Not Found appears
. Click YES.
9.) Confirm the installation by clicking FINISH to close the Hardware Wizard.
10.)Close all open windows, close the “System Setup” with SAVE & EXIT and restart the system
(turn off and on the system).
Figure 3-16 Search for Network path
Figure 3-17 Confirm correct path and finish the Installation
NOTICE
Ver
After boot up of the system, verify the correct settings in the printer “Properties”, see:
Section 3-5-5
"Adjustment of Printer Settings" on page 3-23
.