Section 4-4 - Functional Checks
4-4-7
Basic Measurements
(cont’d)
4-4-7-4
Velocity Measurements (Spectral Doppler Mode)
NOTE:
The Spectral Doppler image is displayed based on time (X-axis) and velocity (Y-axis).
4-4-7-4-1
Acceleration Velocity and Velocity Ratio
1.) Press the
MM
key once.
2.) Select the appropriate item among D VELOCITY or D A/B.
3.) Perform the measurement(s) using the
TRACKBALL
and
SET
(right or left trackball key).
4-4-7-4-2
Average Velocity (Manual Trace)
1.) Press the
MM
key once.
2.) Touch D TRACE on the Touch Panel. A cursor appears on the screen.
3.) Move the cursor with the
TRACKBALL
to the start point of the measurement and press
SET
(right or left trackball key) to fix the marker.
4.) Trace to the end of the period and press the
SET
key again to fix the mark.
The measurement results appear on the screen.
Before you complete the measurement:
To readjust the traced line, press
UNDO
(upper trackball key) repeatedly.
NOTE:
Depending on the setting in the Measure Setup, the envelope curve will be performed with a
continuous trace line or by setting points.
4-4-7-4-3
Average Velocity (Auto Trace)
1.) Press the
MM
key once.
2.) Touch D AUTO TRACE on the Touch Panel.
It traces the Spectral Doppler image automatically and displays the results.
3.) Select the
SENSITIVITY
of the envelope curve (to eliminate artifacts).
4.) Select the
TRACE MODE
channel of the envelope curve (upper, both, lower).
5.) If necessary, select the Angle and the Baseline.
6.) Press the right or left trackball key to
FINISH
the measurement.
Before you complete the measurement:
-
To readjust the start cycle (vertical yellow line), press
CHANGE
(upper trackball key).
Press
SET
(right or left trackball key) to fix the line.
-
Press the
CHANGE
key again to readjust the end cycle (vertical green line).
Press
SET
to fix the line.
NOTE:
The determination of the envelope curve requires a clear and low-noise record of the
Doppler spectrum. Otherwise the reliability of the displayed measurement results may
not be ensured!
Содержание Voluson 730
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL iii ...
Страница 8: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL vi ...
Страница 24: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL xxii ...
Страница 268: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL IV ...
Страница 269: ......