GE M
EDICAL
S
YSTEMS
- K
RETZTECHNIK
U
LTRASOUND
D
IRECTION
105844, R
EVISION
1
V
OLUSON
® 730 S
ERVICE
M
ANUAL
5-14
Section 5-2 - General Information
5-2-4-7
External Video Mode (Video display from VCR)
1.) CPC
Not used for Signal-Processing
2.) CPR
Not used for Signal-Processing
3.) CPG
Not used for Signal-Processing
4.) CPF
Not used for Signal-Processing
5.) CPS
V-Rec-Data from the CKV-Video-Decoder are input to the CPS via the CFM-Data-Bus (CPF must
disable Output-Driver). In the LM-FPGA the data are synchronized and output to the FM-FPGA.
In the Extern-Video Mode the FM-FPGA is specially configured, that’s why the right part of the CPS-
Blockdiagram is not valid in this Mode. In this Mode the FM-FPGA re-scales the Videorecorder-
Image, that is to say that the Video is made for fitting exact into the Monitor-Frame.
6.) CKV
AD-Conversion of the grapped Video, etc.