Section 7-8 - Error Messages
Section 7-8
Error Messages
Error Messages
Actions
530-Probe connected on……….
Disconnect and reconnect the probe. If error remains don't
use such a probe.
Array bounds exceeded
restart the system
Attempt to read Probe-ID from an invalid probe connector.
Reboot system.
AVI Save function fails
check the SCSI and the Power cable - restart the machine
and try again
B_Enhance Out Of Range
Press ok and save this user-setting once again
B_Gain Out Of Range
Press ok and save this user-setting once again
B_Reject Out Of Range
Press ok and save this user-setting once again
B_TxFocus - not calculated and B_SHOT_PART_ON
restart the system, If after restart not ok then call technical
support
BC lines_per_sequenz < 1
restart the system
BC lines_per_sequenz < 2
restart the system
BC_Dynamic Out Of Range
Press ok and save this user-setting once again
BC_Gain Out Of Range
Press ok and save this user-setting once again
BC_Lines: BC_LineDensity out of limit
restart the system
BCMC_Balance Out Of Range
Press ok and save this user-setting once again
BC-TxPower Out Of Range
Press ok and save this user-setting once again
BF: can't set BM RxApod
restart the system, If after restart not ok then call technical
support
BF: can´t set C Rx Apod
restart the system, If after restart not ok then call technical
support
BF: can't set D RxApod
restart the system, If after restart not ok then call technical
support
BM_Resample: overrun SampleLengthOnLineMem
restart the system
Cannot read a valid Probe-ID (%02X) from………….
Disconnect and reconnect the probe
Cannot write Probe-ID into EEPROM on left probe connector
(A).
connect the probe once more, if not ok call technical support
Cannot write Probe-ID into EEPROM on left probe connector
(B).
connect the probe once more, if not ok call technical support
Cannot write RotatingAngleCorr into EEPROM on left probe
connector (A).
connect the probe once more, if not ok call technical support
Cannot write RotatingAngleCorr into EEPROM on left probe
connector (B).
connect the probe once more, if not ok call technical support
Содержание Voluson 730
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL iii ...
Страница 8: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL vi ...
Страница 24: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL xxii ...
Страница 268: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL IV ...
Страница 269: ......