4-10
Section 4-4 - Functional Checks
4-4-2
M-Mode Checks
For further details refer to the Voluson® 730 Basic User Manual, Chapter 6, M-Mode.
Figure 4-8 M Main- and M Sub Menu
Table 4-4
M-Mode Functions
Step
Task
Expected Results
1
Cursor Position
Adjust the M-Cursor position with the
TRACKBALL
in the B-Single image.
2
Activation of M-Mode
Press the
right or left trackball key
to activate both Modes (B/M).
3
M-Mode Gain
Rotate the
M-MODE
key to adjust the sensitivity (brightness) of the entire M-image.
4
M-Mode Depth
Common with B-Mode
Depth
.
5
SPEED
By touching up or down, four different sweep speeds can be selected.
6
INVERT
Invert of the M-Mode image.
7
FREQUENCY
Common with B-Mode
Frequency
.
8
FORMAT
For selection of three different rations of display format.
9
DYN.CONTR.
Dynamic Range enhances a part of the grayscale to make it easier to display pathology.
10
ENHANCE
Due to this function a finer, sharper impression of the image is produced.
11
REJECT
It determines the amplitude-level below which echoes are suppressed (rejected).
Содержание Voluson 730
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL iii ...
Страница 8: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL vi ...
Страница 24: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL xxii ...
Страница 268: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL IV ...
Страница 269: ......