GE M
EDICAL
S
YSTEMS
- K
RETZTECHNIK
U
LTRASOUND
D
IRECTION
105844, R
EVISION
1
V
OLUSON
® 730 S
ERVICE
M
ANUAL
10-20
Section 10-7 - Electrical Safety Tests
10-7-8-1
Data Sheet for ECG Leakage Current
The test passes when all readings measure less than the value shown in the table below. Record all
data on the PM Inspection Certificate.
Table 10-16 Maximum Allowance Limit for ECG Leakage Current
AC Power
Source
Maximum
Allowance Limit
GROUND OPEN
GROUND
CLOSED
Patient Lead to Ground Leakage Current Test
and
Patient Lead to Lead Leakage Current Test
115V
10uA
10uA
220/240V
500uA
10uA
Table 10-17 Maximum Allowance Limit for ECG Leakage Current
AC Power
Source
Maximum
Allowance Limit
Patient Lead Isolation Current Test
115V
20uA
220/240V
5mA
Table 10-18 Typical Data Sheet for ECG Leakage Current
ECG
Power
Tester
Polarity
Switch
Tester
Ground
Switch
Tester Lead Selector
RL
RA
LA
LL
C
ON
NORM
CLOSED
ON
REVERSE
CLOSED
ON
NORM
OPEN
ON
REVERSE
OPEN
OFF
NORM
CLOSED
OFF
REVERSE
CLOSED
OFF
NORM
OPEN
OFF
REVERSE
OPEN
Содержание Voluson 730
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL iii ...
Страница 8: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL vi ...
Страница 24: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL xxii ...
Страница 268: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL IV ...
Страница 269: ......