5-12
Section 5-2 - General Information
5-2-4-3
D-Mode (Pulsed Wave- and Continuous Wave Doppler)
1.) CPC
see:
5-2-4-1 B-Mode
2.) CPR
see:
5-2-4-1 B-Mode
3.) CPG
After DC-cancelling the sig is mixed with the NCO-generated RX-Frequency and brought to LF-
Spectrum, where the FIR-LP cuts HF. Output to CPF.
4.) CPF
Arranges the FFT. D-Mode Data use the same Bus to the CPS as CFM-Data.
5.) CPS
D-Mode Data from CPF are written... see
5-2-4-1 B-Mode
.
6.) CKV
see:
5-2-4-1 B-Mode
5-2-4-4
CFM-Mode (Color Flow Mode)
1.) CPC
see:
5-2-4-1 B-Mode
2.) CPR
see:
5-2-4-1 B-Mode
3.) CPG
After DC-cancelling the sig is mixed with the NCO-generated RX-Frequency and brought to LF-
Spectrum, where the FIR-LP cuts HF. Output to CPF.
4.) CPF
Arranges the CFM-Processing.
WMF (Wall Motion Filter), Autocorrelator, Post-processing.
5.) CPS
Data from CPG and CPF are written... see
5-2-4-1 B-Mode
.
6.) CKV
see:
5-2-4-1 B-Mode
Содержание Voluson 730
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL iii ...
Страница 8: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL vi ...
Страница 24: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL xxii ...
Страница 268: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL IV ...
Страница 269: ......