Section 7-4 - Screen Captures and Logs
Section 7-4
Screen Captures and Logs
There may be times when the customer or field engineer want to capture a presentation on the screen
so it may be recovered by the OnLine Center. This is accomplished by saving the image(s) to Sonoview
and export them to CD-Rom.
7-4-1
Capturing a screen
Only possible if SonoView Option is enabled.
The following is the generic process to capture any screen from the scanner.
1.) Navigate to and display the image/screen to be captures.
2.) Press
SAVE
button and save the image to SonoView.
3.) Select the stored image(s) in SonoView (Exam List) and export the image(s) to CD-Rom (jpg,
bmp or tiff)
7-4-2
Export Log’s and System Data’s
Touch UTILITIES, then SYSTEM SETUP and select the
SERVICE
page.
Type in the password
SHE
and click the ACCEPT button.
7-4-2-1
Export Event Log
Exports Event Log File to MO or CD Drive.
7-4-2-2
Export System Data
Transfer Service Database to MO or CD Drive.
7-4-2-3
Export System Status
This function exports information about probes, boards, SW, Options Service Actions to MO or CD
Drive.
Figure 7-3 Service Tools
Содержание Voluson 730
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL iii ...
Страница 8: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL vi ...
Страница 24: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL xxii ...
Страница 268: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL IV ...
Страница 269: ......