Chapter 7 Diagnostics/Troubleshooting
7-15
7-7-6
Unable to Record to VCR
Figure 7-17 VCR Troubleshooting
Unable to record to VCR!
Check that all signal, and power
supply cable connections to the VCR
unit are OK.
YES
NO
Is the VCR properly
connected?
Is the VCR tape put into
the device?
Put the tape into the device and
rewind it.
NO
YES
Replace VCR.
Is recording impossible even
by manual operation?
YES
- Check the signal cable connections
between VCR and Internal I/O.
NO
YES
Video Output of System
o.k.?
Replace CKV
NO