Chapter 3 Installation
3-35
* RGB OUT has no sync on signals - other values decrease by 300mV
3-6-5
Available Probes
See
Chapter 9 - Probes
,
for part numbers to be used when ordering new or replacement service probes.
3-6-6
Software/Option Configuration
Refer to the Voluson® 730 Basic User Manual, Chapter 17.3.1, System Setup - General, for information
on configuring items like Clinic Name, Language, Display, Date/Time, Date Format and Time Format.
For information on configuring User Settings, refer to Refer to the Voluson® 730 Basic User Manual,
Chapter 17.3.2, System Setup - User Settings.
For information on configuring assignment of
keys, Save Destination, Video norm, etc., refer to
the Voluson® 730 Basic User Manual, Chapter 17.3.3, System Setup - Peripherals.
For information on configuring Software Options, refer to the Voluson® 730 Basic User Manual,
Chapter 17.3.4, System Setup - Options.
For information on configuring DICOM Connectivity, refer to the Voluson® 730 Basic User Manual,
Chapter 17.3.8, System Setup - DICOM.
For information on configuring General Measurement Setup, refer to the Voluson® 730 Basic User
Manual, Chapter 18.3.1.
For information on configuring Obstetric Measurement Setup, refer to the Voluson® 730 Basic User
Manual, Chapter 18.3.2.
For information on configuring Cardiac Measurement Setup, refer to the Voluson® 730 Basic User
Manual, Chapter 18.3.3.
Vertical Front Porch
256us 4 Lines
381us / 6 Lines
Vertical Sync
160us / 2.50 Lines
190us 7 3 Lines
Vertical Back Porch
1408us / 22 Lines
890us / 14 Lines
Serration Pulses
5
6
Interlaced
yes
yes
Aspect Ratio pixel size
14.75 (H) : 17.734475 (V)
14.75 (H) : 14.318 (V)
Video levels on 75 Ohm:
white level
1020mV
1020mV
black level
370mV
370mV
blanking level
320mV
320mV
sync level*
20mV
20mV
Table 3-19 Video Specifications for Composite, B/W, S-Video and RGB Connectors
S-Video Output set to:
Timing Parameter
PAL
50Hz
NTSC
60Hz
Содержание Voluson 730
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL iii ...
Страница 8: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL vi ...
Страница 24: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL xxii ...
Страница 268: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL IV ...
Страница 269: ......