Section 7-5 - How to use the Auto Tester program
5.) Click OK.
6.) Press the
PAUSE/BREAK
key on the keyboard.
7.) Activate the “Auto Tester” program by clicking the “Record” icon on the displayed screen.
8.) Start scanning.
You can scan normally and everything will be recorded to the program (up to several hours.)
NOTE:
It is important that you are recording the processes where the errors normally occur.
9.) Stop the program by pressing the
PAUSE/BREAK
key on the right upper corner of the keyboard.
Figure 7-6 Message Box
Figure 7-7 Start Auto Tester
Pause/Break
key on the
right upper corner
of the keyboard
“Record”
icon