Page 76
Section 9.0 - System Configuration
9.3.3
General Parameters
Table 12-3 General Parameters
Service Contract
Number*
Customer licensed Inhouse Service Contract number; created by GE
Healthcare. (64 characters maximum)
Contract Expiration*
Licensed Inhouse Service Contract number expiration date; created by
GE Healthcare. (64 characters maximum)
AE Title
16 characters maximum
Note: If the calling title is required on the RIS/HIS/PACS,
please configure the same value as this one.
Port Number
4 characters maximum
Parameter
Definition
Parameters
Definition
Time Zone
Allows the user to change/set system time zone. After the time zone
changed, click the “commit” button to apply the changes and then
update the system time.
System Clock
Allows the user to change/set system date and time.
Date Display Format
Allows the user to select date format. The Date format is defined as MM/
DD/YYYY (default) or DD/MM/YYYY.
Time Display Format
Allows the user to select time format. Time Format may be kept in a 24
hour format (hh-mm-ss) or 12 hour AM/PM format: (hh:mm:ss (AM/
PM)). The default, 24 hour format.
Keyboard
Defines the keyboard (English, French, Spanish, Portuguese, Turkish
and Russian) for input.
Country Code
Choose the country appropriate for your installation. The default value is
China.
Measurement System
Used to define English or Metric as the system’s measurement system.
The default value is metric.
Encoding
Three options provided: ISO_IR 100, ISO_IR 192 and GB18030.
GB18030 is only for chinese. ISO_IR 192 is for chinese and
russian. SO_IR 100 is for other languages. If simplified-chinese
input is required, please select GB18030.
Note: Confirm encoding in first installation. Change encoding
during use should cause patient data compatibility error and
cause previous images can be viewed only.
Application Language
Defines the language (English, Chinese, French, Spanish, Portuguese,
Turkish and Russian) used for applications screens. The default value is
English.
QAP Timeout
This item determines the interval between QAP test in days. Used to
inform system users that a QAP interval has expired and a QAP test
should be performed. The default value is 7 days. (Three characters
maximum)
VA Site
This will configure the system for VA options.
Screen Blank Timeout
Used to set the period of inactivity (where no user inputs and processing
activity occurs) in which screen blanking can begin. This feature
functions to prolong the life of the monitor, thus preventing a burn-in.
Time is selected in minutes. The default value is 15 minutes. (4
characters maximum)
Image Mask Adjustment
Adjusts the mask that is applied to the acquired image from the
detector. It affect Raw and Processed images.
Содержание Healthcare Brivo XR118
Страница 4: ...GE HEALTHCARE DIRECTION 5461425 8EN REVISION 6 BRIVO XR118 SERVICE MANUAL Page 4 Legal Notes ...
Страница 46: ...GE HEALTHCARE DIRECTION 5461425 8EN REVISION 6 BRIVO XR118 SERVICE MANUAL Page 46 ...
Страница 134: ...GE HEALTHCARE DIRECTION 5461425 8EN REVISION 6 BRIVO XR118 SERVICE MANUAL Page 134 Section 14 0 Installation Checklist ...
Страница 164: ...GE HEALTHCARE DIRECTION 5461425 8EN REVISION 6 BRIVO XR118 SERVICE MANUAL Page 164 Chapter 7 Brivo XR118 System Check ...