Chapter 10 - - Renewal parts
Page 247
GE H
EALTHCARE
D
IRECTION
5461425-8EN, R
EVISION
B
RIVO
XR118 S
ERVICE
M
ANUAL
3.) Remove old locker from detector and put new locker in the slot.
4.) Push label with linear paper up and put the screw into hole, then tighten the screw with phillips
screwdriver (Torque of Screw: 0.2N·M±0.05N·M).
Put new locker in the slot
After putting the new locker
Push label up