Chapter 8 - - Troubleshooting
Page 219
4.) Click "Continue" button.
14.) Click "Complete" button.
Section 5.0Image Quality Issues
5.1
Equi-spaced Black Bad Line Issue
Although detector calibration/QAP passed, after taking an exposure, there are equi-spaced bad
lines at the top of image. Perform QAP again but failure occurred again because of too many bad
pixels happened in image.