Chapter 10 - - Renewal parts
Portable workstation main board assembly replacement procedure
The device or portions of the device may have power present from the battery even when shut
down, please demount the battery first before this procedure.
1.) Follow the steps of
Portable workstation battery replacement procedure on page 236
to
remove the battery from workstation.
2.) Follow the steps
Portable workstation SSD drive replacement procedure on page 238
to
remove SSD drive.
3.) Follow the steps
Portable workstation RAM replacement procedure on page 240
to remove
RAM.
4.) Demount the screws at the back of the workstation.
5.) Demount the back cover from the workstation.