Chapter 8 - - Troubleshooting
Page 165
Chapter 8 - Troubleshooting
Section 1.0
Hardware Failure
1.1
Using the Built-in Diagnostics tools
Under the Diagnostics tab of SUIF, a set of tools are provided to GEHC service personnel, when
the GEHC restricted service key is installed. Three categories tools are included: Wireless detector,
wireless link and communication.
To utilize these diagnostics tools:
1.) Close the SUIF, if it’s open.
2.) Install the service key CD into the CD-ROM drive and close the drive.
3.) Go to SUIF -> Diagnostics.
4.) Using the mouse, left click on the tool interest to you.
5.) Follow the instructions on the page to conduct the self diagnostic.
6.) To exit, click on the SUIF <close >button and remove the service key CD.