Page 128
Section 12.0 - Leakage Current Test
12.1.1
Personnel Requirements
12.1.2
Overview
This procedure can be performed with:
• Fluke tester
12.1.3
Tools and Test Equipment
Note:
The equivalent means that the tool is designed to be capable of performing tests to both the
standard of IEC 62353 and the earth leakage current of IEC 60601-1.
RISK OF ELECTRIC SHOCK.
VOLTAGE PRESENT.
THE SYSTEM MUST BE DISCONNECTED FROM THE FACILITY MAINS INCLUDING THE A1
PANEL GROUND WIRE.
12.1.4
Setup the tools
Refer to Chapter 1
12.1.5
Leakage Current Test Procedure
• Earth Leakage Current Test
1) Connected the BRIVO XR118 power plug to tool output receptacle.
Personnel Requirements
Preliminary Reqs
Procedure
Finalization
1
Not Applicable
30min
10min
Item
Qty.
Effectivity
Part#
Manufacturer
Fluke ESA612 or equivalent
1
-
-
Fluke
No
TEST POINT
Expected Value
(mA)
Measure Value
(mA)
RESULT
1
Connected the BRIVO XR118
power plug to tool output
receptacle
10
Содержание Healthcare Brivo XR118
Страница 4: ...GE HEALTHCARE DIRECTION 5461425 8EN REVISION 6 BRIVO XR118 SERVICE MANUAL Page 4 Legal Notes ...
Страница 46: ...GE HEALTHCARE DIRECTION 5461425 8EN REVISION 6 BRIVO XR118 SERVICE MANUAL Page 46 ...
Страница 134: ...GE HEALTHCARE DIRECTION 5461425 8EN REVISION 6 BRIVO XR118 SERVICE MANUAL Page 134 Section 14 0 Installation Checklist ...
Страница 164: ...GE HEALTHCARE DIRECTION 5461425 8EN REVISION 6 BRIVO XR118 SERVICE MANUAL Page 164 Chapter 7 Brivo XR118 System Check ...