Section 6.0 - Detector Installation
6.4
Detector Placement
Note:
More detailed information about detector placement, please refer to the “Digital Detector
Application” chapter in Operator Manual.
6.4.1
Table upgrade
1.) Remove the film cassette from the table housing if any.
2.) Insert the image detector into the table housing.
6.4.2
Wall stand upgrade
1.) Remove the film cassette from the wallstand housing if any.
2.) Insert the image detector into the wallstand housing.
Brivo XR118 detector is about 3.5kg which is heavier than casette used in analog system. After
the upgrade, balance weight for wallstand cassette housing is required.
6.4.3
Detector Operation
Wake up
Press and hold the power button for more than 2 seconds to wake up the detector when the detector
is in sleep mode.
Sleep Mode
Press and hold the power button for more than 4 seconds to put the image detector into sleep mode
when detector is not in sleep mode.
LED Indicator