GE H
EALTHCARE
D
IRECTION
5461425-8EN, R
EVISION
B
RIVO
XR118 S
ERVICE
M
ANUAL
Page 170
-
1.4.2
Workstation stops working after unplugged from dock
1.4.3
Workstation battery life is short
Unplug workstation, workstation
stops wo rking
Connect workstation with power
Adapter, power on workstation
Workstation boots up?
N
Y
Replace workstatio n battery
Replace mainboard
Workstation battery life is short
Use after completely
charged?
N
Y
Replace mainboard
Please charge the workstation before
the usage
Charge the workstation with dock
Boot the system and enter into
battery diagnostic page
A hint for battery
replacement?
Backup and re-install system
N
Replace workstatio n battery
Problem resolved?
Problem resolved?
Y
N
N