Section 8.0 - Software Installation
8.3
System Software Installation
8.3.1
Personnel Requirements
8.3.2
Overview
•
The following times are typical for a complete system LFC:
1.) Five Minutes to backup system files to CD/DVD.
2.) Approximately ninety minutes to save images and protocols.
3.) Fifteen minutes to install the OS and Applications software.
4.) Five Minutes to restore system files from CD/DVD.
5.) Five minutes to cleanup and return system to operational readiness.
•
5461425-8EN
Only use this System Installation procedure to re-install and/or upgrade your system software
completely. The following procedure installs both the operating system, and application
software.
•
Pay particular attention, if instructed, to system resets. It's important that power be cycled
according to a specified procedure. Cycling power at the wrong time or by the wrong method
can result in a PC's operating file system (Linux) becoming corrupted. If the OS on the PC gets
corrupted, you may have to reload it again or start over.
•
The following SW procedure assumes that you are beginning with the System powered “ON”
and application software running (i.e. System Operationally Ready).
Note:
Configure the Time Zone after software re-installation or upgrade.
Required
Persons
Preliminary
Reqs
Procedure
Finalizati
on
1
5 mins
15 (Time estimate does not include saving
and restoring patient images and/or
protocols files)
0 min
Содержание Healthcare Brivo XR118
Страница 4: ...GE HEALTHCARE DIRECTION 5461425 8EN REVISION 6 BRIVO XR118 SERVICE MANUAL Page 4 Legal Notes ...
Страница 46: ...GE HEALTHCARE DIRECTION 5461425 8EN REVISION 6 BRIVO XR118 SERVICE MANUAL Page 46 ...
Страница 134: ...GE HEALTHCARE DIRECTION 5461425 8EN REVISION 6 BRIVO XR118 SERVICE MANUAL Page 134 Section 14 0 Installation Checklist ...
Страница 164: ...GE HEALTHCARE DIRECTION 5461425 8EN REVISION 6 BRIVO XR118 SERVICE MANUAL Page 164 Chapter 7 Brivo XR118 System Check ...