Page 100
Section 9.0 - System Configuration
9.11.6
Fuji DryPix 7000 Printer Parameters
Table 12-20 Fuji DryPix 7000 Printer Parameters
TAB
Attribute
Value
Comments
Printer
DICOM Printer Label
DRYPIX7000
Or customers preference
Application Entry Title
Enter value given by Site Network Administrator
AE Titles may be site
specific, be sure to check
on-site
IP Address
Enter value given by Site Network Administrator
Port Number
104
Port numbers may be site
specific, be sure to check
on-site
Pixel Depth
12
Printer Pixel Size (micron)
50
Configuration Information
CS000
Density
Min. = 20, Max. = 300
Magnification Type
CUBIC
Smooth Factor
SHARP
Trim
NO
Not Supported
Polarity
NORMAL
Border Density
BLACK
Empty Image Density
BLACK
Printer Memory Size
48Mbytes
Layout
Select the desired layouts
allowed for this printer
1:1, 1:2, 2:1, 2:2
Slide formats