Section 3.0
Wireless Architecture
1.) Using scenario for Fix RAD room with monitor and AP.
2.) Wireless SSID and security configuration
•
Detector’s default IP is fixed
to
3.1.1.30/24.
•
AP’s default IP is fixed to
3.1.1.1/24.
•
ETH1’s default IP on iServer to
AP is fixed to 3.1.1.
/24.
•
ETH2’s IP on iServer to
hospital
LAN is assigned by
hospital
network.
•
SUIF provide an access for FE
to modify iDR subnet(3.1.x.x)
in
case iDR subnet conflicting
with hospital network.
On fixed RAD room
•
AP’s SSID=Associate detector’s MAC
•
AP’s SSID Encryption=WPA2-PSK
•
AP’s SSID Passphrase=Encryption algorithm(detector’s
MAC)
•
AP’s SSID broadcast=disabled
Wifi Connection
•
On fixed RAD room, detector will connect to AP.