GE H
EALTHCARE
D
IRECTION
5461425-8EN, R
EVISION
B
RIVO
XR118 S
ERVICE
M
ANUAL
Page 166
-
1.2
Using the Enhanced Error Messages
Enhanced error messages are available to GEHC personnel, when the GEHC restricted service key
is installed. The enhanced error messages contain additional information not found in the system
error log during normal system operation or available to the normal user. These enhanced error
messages contain additional information that aid in fault isolation and choosing the proper
corrective action.
With the service key installed, error messages in the system error log area automatically linked to
the enhanced messages. The complete set of enhanced error messages is contained in a single
HTML file located on the same CD-ROM as the service key. When the CD is installed, you'll notice
that error numbers in the system log are underlined to indicate a hyper-link to the enhanced
message. You use the mouse to click on the links.
To utilize the enhanced error messages:
1.) Close the SUIF, if it’s open. Applications can only mount the CD-ROM path in Linux, when the
SUIF is closed.
2.) Install the service key CD-ROM into the CD-ROM drive and close the drive.
3.) Go to SUIF -> Error Logs.
4.) Using the mouse, left click on an error number of interest to you. This will launch a browser
and allow you to see the enhanced error message.
5.) To exit, click on the SUIF <close >button and remove the service key CD-ROM.
Содержание Healthcare Brivo XR118
Страница 4: ...GE HEALTHCARE DIRECTION 5461425 8EN REVISION 6 BRIVO XR118 SERVICE MANUAL Page 4 Legal Notes ...
Страница 46: ...GE HEALTHCARE DIRECTION 5461425 8EN REVISION 6 BRIVO XR118 SERVICE MANUAL Page 46 ...
Страница 134: ...GE HEALTHCARE DIRECTION 5461425 8EN REVISION 6 BRIVO XR118 SERVICE MANUAL Page 134 Section 14 0 Installation Checklist ...
Страница 164: ...GE HEALTHCARE DIRECTION 5461425 8EN REVISION 6 BRIVO XR118 SERVICE MANUAL Page 164 Chapter 7 Brivo XR118 System Check ...