Page 246
Chapter 10 - - Renewal parts
Detector Battery Locker Assembly Replacement
Material/Tools/SW/systems needed:
•
Phillips screwdriver
Note:
Below steps show how to replace battery left locker assembly, replacement of the right locker
assembly follows the same steps.
1.) Remove battery from detector.
2.) Use phillips screwdriver to penetrate the label in specified area to expose the screws, then
loose the screws.
Penetrate here
Loose the exposed screws