Chapter 5 - - Installation & Configuration
Page 71
8.4
Restore
8.4.1
Before You Begin
Note:
The detector calibration data table_IDC can not be restored in different system.
Note:
If you wish to use a USB flash drive, select [USB DRIVE] and install drive into USB port.
•
After the OS Configuration restore is run, the SUIF will no longer respond because of the files
that have been changed. A system Reset must be performed after restoring the OS
Configuration files to continue using the SUIF.
•
You must have a “good” backup of the files you wish to restore from CD/DVD.
•
The “Backup” media can only be inserted after the restore utility screen is displayed.
8.4.2
Restore System Files
1.) From the application desktop, click on the Utilities icon.
2.) Click on [Launch] button.
3.) Select the Utilities Tab,
4.) Click on the plus sign preceding the System folder ICON, in the navigation frame.
5.) Click on [Restore] link,
6.) Insert your "Cal/Config Backup" CD/DVD into the DVD drive.