Chapter 1 - - Safety and Service
Page 25
Section 2.0
Applying System Power
2.1
System Preparation for Energizing Power
POTENTIAL
ELECTRICAL
SHOCK
PERFORM LOTO BEFORE PERFORMING SERVICE OF ANY KIND.
Follow the steps bel ow to ensure that your system powers up safely, properly and reliably.
Potential for
Equipment
Damage
Cabinet settings must match input voltage else there is potential for equipment damage.
The power cord plug complies with the local national standard.
2.2
Energizing System
The Brivo XR118 system is designed to function totally independent of the X-ray RAD target system
without any hardware connection. So its power-up is controlled by connected power socket.
There is no recommended boot up sequence for XR118 system and target system.
Complete XR118 system startup takes approximately 3 minutes.