GE H
EALTHCARE
D
IRECTION
FQ091013, R
EVISION
1
V
IVID
I
N
AND
V
IVID
Q
N S
ERVICE
M
ANUAL
8-136
Section 8-5 - Software Loading
9) Run
PMCbootloader.exe
by double-clicking on it - see
Figure 8-132 Run PMCbootloader.exe
A window is displayed showing the current PMC version and the path of the PMCJPS2.hex file:
10.)Click
Burn
.
The burning process will take approximately 30 seconds.
When done, a brief message will appear “
Burning process failed…
.
Note:
The failed message is normal and should be anticipated.
Figure 8-133 Burning Process Failed Message