Chapter 4 - General Procedures and Functional Checks
4-29
4-3-2-2
B/W and Color Printer Test
4-3-2-3
Color Printer Test
Table 4-17 B/W Printer Test - Recommended
Step
Task
Expected Result(s)
1
Scan a phantom or suitable interface in 2D-Mode.
2
Press the
Freeze
button.
The image scanning stops with the latest picture frozen
on the screen.
3
Press the
button.
Evaluate the print quality. Adjust the
brightness
and/or
contrast
of the B/W printer, if required. The image
displayed on the screen is printed on the B/W or color
printer, depending upon the key assignment
configuration.
Table 4-18 Color Printer Test - Recommended
Step
Task
Expected Result(s)
1
Scan a blood vessel in 2D-Mode or a phantom
simulating a similar CFM effect, then activate CFM to
see color.
2
Press the
Freeze
button.
3
Press the
Print P1
button.
Evaluate the color print quality. Adjust the
brightness
and/or
contrast
of the Color printer, if required.