GE H
EALTHCARE
D
IRECTION
FQ091013, R
EVISION
1
V
IVID
I
N
AND
V
IVID
Q
N S
ERVICE
M
ANUAL
Chapter 4 - General Procedures and Functional Checks
4-27
4-3-1-9
Front End Test
4-3-1-10
CW Test
Table 4-13 Front End Test - Recommended
Step
Task
Expected Result(s)
1
Boot up the system in Maintenance Mode, using the
Service Dongle.
2
Press
Alt + Config
to enter Diagnostics Mode
3
From the Diagnostic Test tree, select
Front End Tes
t.
4
Run
Full System Test
and press
Start
PASS
5
Connect the 3S-RS probe to the probe port.
6
From the Diagnostic Test tree, select
Front End Test
.
7
Select
TR32
then select
TX
. Select
TXProbeTX Tes
t
and press
Start
PASS
8
Select
MUX Test
then select
3S-RS Test
and press
Start
PASS
9
Select
TR32 Test
then select
RX Tes
t. Select
RSTest
with 3S-RS
and press
Start
PASS
10
When done, press the
Esc
button
Table 4-14 CW Test - Recommended
Step
Task
Expected Result(s)
1
Boot-up the System in Regular Scanning Mode
2
Press
Alt + Config
to enter Diagnostics Mode
3
From the Diagnostic Test tree, select
Front End Test
.
4
Run
Full System Test
and press
Start
PASS
5
Connect the 3S-RS probe to the probe port.
6
From the Diagnostic Test tree, select
Front End Tes
t.
7
Select
TR32
then select
TX
. Select
TXProbeTX Test
and press
Start
PASS
8
Select
MUX Test
then select
3S-RS Test
and press
Start
PASS
9
Select
TR32 Test
then select
RX Tes
t. Select
RSTest
with 3S-RS
and press
Start
PASS
10
When done, press the
Esc
button