Chapter 8 - Replacement Procedures
8-133
When done, the Installation Complete message is displayed:
10) Move the cursor over the displayed message window and press the
SET
button.
11.)Press any key to re-boot the system.
The system re-boots into Normal Scanning mode.
12.)Press
Config
and then
About
.
The newly-installed software version is now listed under the System Version tab:
13.)Remove the DVD from the
DVD\CD-RW
drive and close the tray.
14.)Proceed to perform the following functionality tests:
-
Software Configuration Checks on page 4 - 37
Figure 8-129 Software Installation Complete Message