Section 3-7 - Completing the Hardware Installation
3-7-2-5
External USB Hard Disk
The External USB Hard Disk can be connected to any of the USB ports on the SafeLock Cart.
One must use the additional power source (supplied with the External USB Hard Disk) to activate the
External USB Hard Disk.
WARNING
Do not attempt to use a different type of mass storage device other than the
External USB Hard Disk supplied with the system. The ultrasound system is an extremely
sensitive and complex medical system. Any unauthorized peripherals may cause system failure
or damage!
Refer to the
External USB Hard Disk Installation Procedure
section, page 8-151 for instructions on initial
configuration and connecting the External USB Hard Disk to the Vivid
i
n/Vivid q N portable ultrasound
scanner.
NOTICE
IMPORTANT
Do not connect the External USB Hard Disk to the Vivid i
n/ Vivid q N
system while
scanning is in progress! This will avoid the risk of causing image artifacts.
NOTICE
IMPORTANT
During Stand-by mode, it is NOT recommended to
introduce
or
remove
USB
devices; this may cause the system to lock-up during the boot-up procedure
.
Figure 3-39 External USB Hard Disk