Section 3-4 - Unpacking the Equipment
3-4-3
Verifying the Transportation box Contents
After unpacking, it is important to verify that all items ordered by the customer have been received (refer
to the Delivery Note). Compare all items listed on the packing slip (shipping consignment note) with
those received and report any items that are missing, back-ordered, or damaged, to your GE Healthcare
sales representative.
3-4-4
Physical Inspection
3-4-4-1
System Voltage Settings
Verify that the Vivid
i
n/ Vivid q N AC adapter is set to the correct voltage. The Voltage settings are
220-240V AC.
3-4-5
EMI Protection
The Vivid
i
n/ Vivid q N Ultrasound Unit has been designed to minimize the effects of Electro Magnetic
Interference (EMI). Many of the covers, shields, and screws are provided primarily to protect the system
from image artifacts caused by this interference. For this reason, it is imperative that all covers and
hardware are installed and secured before the unit is put into operation.
NOTICE
It is recommended to keep and store the transportation box and all other packing materials in
case the unit has to be moved to a different location in the future.
For warranty purposes, storage of the above is required for one year from date of purchase.
WARNING
WARNING
Setting the
ultrasound components
to the wrong voltage setting will most likely destroy the
equipment.