GE H
EALTHCARE
D
IRECTION
FQ091013, R
EVISION
1
V
IVID
I
N
AND
V
IVID
Q
N S
ERVICE
M
ANUAL
Chapter 4 - General Procedures and Functional Checks
4-25
4-3-1-8
ECG/Respiratory Test
10
Press
Alt +Config
to start the diagnostic package,
11
From the diagnostic test tree, select
BEP
.
This will activate PC Doctor diagnostics.
12
Run
System Diagnostics\Monitor Test
13
Select
Combination Test
.
14
Follow the on-screen instructions.
PASS
15
When done, press
Close
The system will prompt that the test passed.
Table 4-12 ECG/Respiratory Test - Recommended
Step
Task
Expected Result(s)
1
Boot up the system in Maintenance Mode, using the
Service Dongle and load the application by
double-clicking the
Echoloader
icon on the desktop.
2
Connect the ECG cable to the input on the ECG port on
the right side of the Vivid
i
n/ Vivid q N scanner.
Connect a Cardiac probe (for example, 3S-RS) and
press on
Physio
. Set ECG as OFF and Int/Resp as ON.
Connect the ECG leads to yourself (as the simulated
patient). Refer to the instructions in the Vivid
i
n/Vivid q N
User Manual.
3
Press
Alt + Config
to enter into Diagnostics Mode
4
From the Diagnostic Test tree, select
BEP
.
5
Run the
ECG Test
6
From the Test ECG device window, set the resolution to
1 second and press the
Open Device
button
7
Verify that a single clear QRS signal is displayed.
Table 4-11 Monitor Test (cont’d) - Recommended
Step
Task
Expected Result(s)
Содержание FL000090
Страница 2: ......
Страница 30: ...GE HEALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL xxviii ...
Страница 739: ......