GE H
EALTHCARE
D
IRECTION
FQ091013, R
EVISION
1
V
IVID
I
N
AND
V
IVID
Q
N S
ERVICE
M
ANUAL
8-104
Section 8-5 - Software Loading
8-5-1-2
Vivid
i
n/ Vivid q N Software Re-Installation Procedure
NOTE:
The following software re-installation procedure is based on the Vivid i n system. The procedure
is the same as for the Vivid q N system.
Table 8-6
Vivid
i
n/ Vivid q N Software Re-Installation
Step
Instruction
Expected Result
1.
Remove the battery and remove all peripherals currently
connected to the system, with the following exceptions:
Make sure no peripherals are connected to the USB
connectors except the DVD and the PCMCIA.
2.
Press the <Eject> button of the DVD\CD-RW drive and
remove any media if present.
3.
Insert the CD provided with the software kit into the
DVD\CD-RW.
The DVD\CD-RW drive loads the software CD.
4.
Hold down the <ON/OFF> button for 5 seconds to
shutdown the system.
The shutdown options are displayed on-screen.
5.
Select
Full Shutdown
.
The system powers down.
6.
Hold down the <ON/OFF> button for three seconds.
The system powers ON and when system boot-up from the CD is complete, the
following message is displayed prompting you to select the required option:
7.
Select option 1.
Vivid_i Full Clean Install ( Application
Software
)
Follow the on-screen instructions, then wait until the following message is
displayed, prompting you to select the required option:
8.
Select option A.
Clean "Vivid i" Install System S/W ...
The user is prompted for a password.