Chapter 8 - Replacement Procedures
8-71
8-4-7
Lower Cover (Rear) Replacement Procedure
8-4-7-1
Tools
Appropriate Phillips screwdriver.
8-4-7-2
Preparations
Shut down the Vivid
i
n/ Vivid q N ultrasound unit, as described in
8-4-7-3
Lower Cover (Rear) Removal Procedure
Remove six retaining screws of the lower cover (rear) and while supporting the cover with one hand,
gently remove the cover - see
8-4-7-4
Lower Cover (Rear) Installation Procedure
1) Return the cover to the correct position on the rear of the SafeLock Cart - see
.
2) While holding the cover to prevent it from falling, fasten the retaining screws of the lower cover
(rear).
3) Proceed to perform the following functionality tests:
-
SafeLock Cart - Grounding Continuity on page 10 - 23
-
SafeLock Cart - Chassis Current Leakage Test on page 10 - 25
Figure 8-82 Lower Cover (Rear) Removal
Retaining Screw