GE H
EALTHCARE
D
IRECTION
FQ091013, R
EVISION
1
V
IVID
I
N
AND
V
IVID
Q
N S
ERVICE
M
ANUAL
5-44
Section 5-9 - ECG/Respiratory Module
5-9-2-2
Specifications
5-9-2-3
Connectors
Table 5-17 External ECG/Respiratory Interface Specifications
Characteristic
Specification
Input Range (ac)
2v peak to peak
A2D resolution
12 bit ,2'complement
Sample Rate
500 [samples/second]
Ground Isolation
4 kV rms
Table 5-18 Connector Pin Out J5 (ECG, Respiration, External ECG)
Pin Number
Specification
1
External ECG Input
2
ECG Input, Respiration Input/Output (LL)
3
ECG Input, Respiration Input/Output (RA)
4
ECG Respiration Output (LA)
5
External ECG Select Signal Input (Short to GND on Cable)
6
GND (GND for External ECG or Shield for ECG)
1
3
2
4
5
6