Section 8-2 - External Component Replacement Procedures
GE MEDICAL SYSTEMS FOR PROPER DISPOSAL.
8-2-2-5
Battery Installation Procedure
WARNING: The Vivid
i
n/ Vivid q N
battery has a safety device.
DO NOT ATTEMPT TO DIS-ASSEMBLE OR ALTER THE BATTERY
!
Always observe the precautions listed in
Vivid i n/ Vivid q N Battery Safety on page 1 - 17
CAUTION:
USE ONLY BATTERIES APPROVED BY GE MEDICAL SYSTEMS AS
SUITABLE FOR USE WITH THE VIVID
I
n/Vivid q N portable ultrasound scanner
1) With the Vivid
i
n/ Vivid q N lying face down (refer to
), hold the unit steady
with one hand while sliding the battery sideways into the battery compartment.
2.) Make sure the battery is properly inserted all the way, before pushing the locking latch to the closed
position.
3.) Turn the Vivid
i
n/ Vivid q N over, release the front latches and open the LCD cover.
4.) Reconnect the AC adapter and all cables to the Vivid
i
n/Vivid q N portable ultrasound scanner.
5.) Proceed to perform the following functionality tests:
-
SafeLock Cart - Grounding Continuity on page 10 - 23
- only if the Vivid
i
n/ Vivid q N system
is currently mounted on the SafeLock cart
-
SafeLock Cart - Chassis Current Leakage Test on page 10 - 25
-
Calibration Tests on page 7 - 25
-