Chapter 3 - System Setup
3-55
3-7-3
Charging the Battery
NOTE:
The
CH5000GE
is a stand-alone desktop smart-battery charger with the added ability to recalibrate the
fuel gauge when detected to be out of calibration. The charger is specifically designed to operate with
the Vivid i n/ Vivid q Nsmart-battery (P/N 2378963-2).
The charger is supplied in a carton containing the following items:
•
CH5000 desktop charger/calibrator
•
24V 2.5A AC-adapter power supply, universal mains input
•
Instruction Manual
CAUTION: Always observe the following safety precautions:
•
Do not expose the charger or power supply to water or conductive liquids (this is not a sealed case).
•
Do not open the charger or power supply case. There are no user-serviceable parts inside.
•
Do not cover the fan exhaust or obstruct the airflow, as this will cause overheating.
•
Use only the manufacturer’s 24V 2.5A power supply.
•
Place the charger in a cool location, away from external heat sources.
•
During recalibration, the battery connector and base of the charger may become warm.
3-7-3-1
Setting up the Charger
1) Place the charger unit on a flat, level surface away from sources of heat and moisture.
2) Plug the DC connector from the power supply into the back of the charger.
3.) Connect the power supply to the mains AC supply, using the cable supplied.
Figure 3-56 Battery Charger Unit