Chapter 8 - Replacement Procedures
8-135
8-5-9
Installation of PMC Version
1.) Allow the system to boot up into
Normal Scanning
mode.
2) Press
Config
.
The system will prompt you to login.
3) Login as
ADM
.
The default password is
ulsadm
4) Press the
About
button and activate the
FwVersion
tab; verify that PMC version is 19 or above, as
shown in the example in
5) Turn the system OFF and remove the DVD USB cable and both battery and AC adapters.
6) Connect the AC adapter
only
(NO BATTERY).
7) Boot up the system in
Maintenance
mode, using the Service Dongle.
8) Go to the following directory: My Computer
C:\marlin\Target\bin
.
Figure 8-131 Verifying PMC Version