GE H
EALTHCARE
D
IRECTION
FQ091013, R
EVISION
1
V
IVID
I
N
AND
V
IVID
Q
N S
ERVICE
M
ANUAL
3-34
Section 3-7 - Completing the Hardware Installation
3-7-2
Connecting Peripherals
NOTE:
All peripherals mounted on the SafeLock Cart must be firmly secured in position using the securing
straps supplied with the SafeLock Cart. These straps are fastened through the metal bars located on
each side of the cart and serve to hold the peripheral firmly in place.
For instructions on mounting the Vivid i n/ Vivid q N scanner on the SafeLock Cart, refer to
the Vivid i n/ Vivid q N on the SafeLock Cart (optional) on page 3-65
Following the connection and removal of peripherals, perform the following functionality tests:
•
SafeLock Cart - Grounding Continuity on page 10-23
•
SafeLock Cart - Chassis Current Leakage Test on page 10-25
3-7-2-1
USB Global Modem
The USB Global Modem connection in the different scenarios is explained below.
•
No SafeLock Cart - connection directly to the Vivid
i
n/ Vivid q N
system
The USB Global modem can be powered via the USB port on the Vivid
i
n/ Vivid q N system with no
need for additional power. The modem should be connected to the
lower
USB port on the rear
connectors panel.
When connecting the modem, there is no preference as to which cable to connect first. One end of the
telephone cable can be connected to the wall outlet and the other side to the modem, followed by
connecting one end of the USB cable to the modem and the other side to the
lower
USB port on the
rear connectors panel.
All software drivers for the modem are pre-installed for the specific modem type. One must follow the
i-Linq configuration procedure to complete the check-out process.
WARNING
Do not attempt to use a different type of modem either via the PC MCIA or USB ports.
The ultrasound system is an extremely sensitive and complex medical system. Any
unauthorized peripherals may cause system failure or damage!
•
No SafeLock Cart - connection via USB hub
This is not an approved configuration and, in most cases, will not work and will cause software
complications.
•
Vivid
i
n/ Vivid q N mounted on SafeLock Cart
This is not an approved configuration and, in most cases, will not work and will cause software
complications.
NOTICE
IMPORTANT
During Stand-by mode, it is NOT recommended to
introduce
or
remove
USB
devices; this may cause the system to lock-up during the boot-up procedure
.