GE H
EALTHCARE
D
IRECTION
FQ091013, R
EVISION
1
V
IVID
I
N
AND
V
IVID
Q
N S
ERVICE
M
ANUAL
3-66
Section 3-8 - Mounting the Vivid i n/ Vivid q N on the SafeLock Cart (optional)
3-8-1 Mounting the Vivid i n and Vivid q N on the SafeLock Cart
2.) Push the scanner down and
backwards
, until it locks into place against the locking tooth.
An audible click will be heard as the scanner locks into position. There will now be a gap the width
of one finger between the SafeLock Cart arm rest and the front of the scanner.
3.) If necessary, perform the alignment procedure as described in
3-8-2 "Adjusting the SafeLock Cart
Locking Mechanism" on page 3-68
.
4.) Connect the LAN Cable from the cart to the LAN connector on the Vivid
i
n and Vivid q N rear panel.
5.) Connect the Docking Connector to the rear of the scanner, as shown below.
Figure 3-63 Vivid
i
n and Vivid q N Scanner Mounted on SafeLock Cart
Figure 3-64 Vivid
i
n and Vivid q N
Rear Panel
Figure 3-65 Connecting the Docking Connector to the Rear of Scanner
Docking Connector
LAN Connector
USB Port
Docking Connector
Содержание FL000090