Chapter 3 - System Setup
3-3
3-2-3
Safety Reminders
Important
: When the SafeLock Cart is connected to the wall outlet and the main circuit breaker is in the
ON position, the AC Box Fan is operated, even if the system is turned OFF.
Note:
The Vivid
i
n/Vivid q N User Manual
should be fully read and understood before operating the
unit. Keep the manual near the unit for reference.
DANGER
WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE
AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DO NOT TOUCH
THE UNIT!
WARNING
WARNING
Two people are required to unpack the SafeLock Cart, as it is heavy. Two people are always
required whenever a part weighing 19kg (35 lb.) or more must be lifted.
CAUTION
If the unit is very cold or hot, do NOT turn ON power to the unit until it has had sufficient time to
acclimate to its operating environment.
CAUTION
To prevent electrical shock, connect the unit to a properly grounded power outlet.
Do NOT use a three-prong to two-prong adapter, as this defeats safety grounding.
CAUTION
To ensure proper grounding, connect this equipment to a receptacle marked "HOSPITAL ONLY"
OR "HOSPITAL GRADE".
CAUTION
Do NOT wear the ESD wrist strap when you work on live circuits where more than
30 V peak is present.
CAUTION
Do NOT operate the unit unless all board covers and frame panels are securely in place, to
ensure optimal system performance and cooling. (When covers are removed, EMI may be
present).
CAUTION
ACOUSTIC OUTPUT HAZARD
ALTHOUGH THE ULTRASOUND ENERGY TRANSMITTED FROM THE VIVID
I
N AND VIVID Q N
PROBE IS WITHIN AIUM/NEMA STANDARDS, AVOID UNNECESSARY EXPOSURE.
ULTRASOUND ENERGY CAN PRODUCE HEAT AND MECHANICAL DAMAGE.