Revision B
120 Series Maternal/Fetal Monitor
9-29
2015590-001
General Maintenance: Maternal Blood Pressure Tests
F
pass
F
fail
US:
Connect an US test body to the monitor’s front panel
US
input. The US
test body should be an ultrasound transducer wrapped in aluminum foil.
Measure the voltage breakdown.
F
pass
F
fail
US2:
Connect an US test body to the monitor’s front panel
US2
input. The US
test body should be an ultrasound transducer wrapped in aluminum foil.
Measure the voltage breakdown.
F
pass
F
fail
Mains–to–Chassis Using DC Voltage for One Minute
Connect the monitor to the hi-pot tester as shown in
Figure 9-7
.
NOTE:
The hi-pot tester voltage is 2.121 kVdc.
F
pass
F
fail
Figure 9-7. Mains–to–Chassis Using DC Voltage for One Minute (2.121 kVdc)
Содержание Corometrics 126
Страница 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 2: ......
Страница 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Страница 30: ...1 10 120 Series Maternal Fetal Monitor Revision B 2015590 001 Safety Equipment Symbols For your notes ...
Страница 404: ...A 6 120 Series Maternal Fetal Monitor Revision B 2015590 001 Factory Defaults Table of Defaults For your notes ...
Страница 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Страница 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Страница 412: ......
Страница 414: ......
Страница 416: ......
Страница 418: ......
Страница 420: ......
Страница 422: ......
Страница 424: ......
Страница 426: ......
Страница 428: ......
Страница 430: ......
Страница 432: ......
Страница 434: ......
Страница 436: ......
Страница 438: ......
Страница 439: ......