8-14
120 Series Maternal/Fetal Monitor
Revision B
2015590-001
Self-Tests: Diagnostic Control Screen
Corolan Network Status
If Corolan Address Checking is
disabled
on the Install Options screen, this field will
normally display NO ADDRESS. If Corolan Address Checking is
enabled
on the
Install Options screen, the monitor periodically performs an address test on the
Corolan wallplate the monitor is attached to. If the wallplate address lines are intact,
this field will display an OK status. If the monitor detects a problem, the monitor
automatically goes offline from the Corolan network.
Table 8-3
provides an
explanation of the more common status indications that can be listed in this field.
Table 8-2. Main Board SW1 Switch Settings
Switch #
Description
On (0) Position
Off (1) Position
State
Display
Notation
State
Display
Notation
8
MECG Option
Disabled
Enabled
MECG
7
MSpO
2
Option
Disabled
Enabled
MSPO2
6
NBP Option
Disabled
Enabled
NBP
5
FSpO
2
Option
Disabled
Enabled
FSPO2
4
Clock Speed
16 MHz
16 MHz
20 MHz
3
Unused
—
—
2
J102 Output
Levels
Corometrics
CORO
Hewlett-
Packard
1
Factory Test
Disabled
Enabled
TEST
Table 8-3. Corolan Network Status Messages
Message
Meaning
OK
Monitor is successfully communicating with the Corolan master device (e.g.
QS Server).
JUST STARTED
Contact GE Service Representative.
ADDRESS NEVER CHECKED
FLOATING ADDRESS
There is a faulty address line in the wallplate.
INVALID ADDRESS CHANGE
A wallplate has been changed from one address to another. This could be
the result of a partially disconnected cable.
ADDRESS KEEPS CHANGING
Contact GE Service Representative.
ADDRESS CHANGING
Содержание Corometrics 126
Страница 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 2: ......
Страница 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Страница 30: ...1 10 120 Series Maternal Fetal Monitor Revision B 2015590 001 Safety Equipment Symbols For your notes ...
Страница 404: ...A 6 120 Series Maternal Fetal Monitor Revision B 2015590 001 Factory Defaults Table of Defaults For your notes ...
Страница 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Страница 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Страница 412: ......
Страница 414: ......
Страница 416: ......
Страница 418: ......
Страница 420: ......
Страница 422: ......
Страница 424: ......
Страница 426: ......
Страница 428: ......
Страница 430: ......
Страница 432: ......
Страница 434: ......
Страница 436: ......
Страница 438: ......
Страница 439: ......