9-14
120 Series Maternal/Fetal Monitor
Revision B
2015590-001
General Maintenance: Maternal Blood Pressure Tests
Patient-to-Ground Leakage for ECG
1.
Connect an ECG test body to the monitor’s front panel
ECG
input. The ECG test
body can be shorted MECG cable, or the assembly leakage cable.
IMPORTANT
TEST CABLES—If you use transducers in lieu of the assembly
leakage cable, and you are testing a Model 129/129F Monitor, you
must test both FECG and MECG.
2.
Set the fixture’s
Leakage
switch to
Patient
.
3.
Turn ON the monitor using the monitor’s power switch.
4.
Verify the following:
F
< 50
µ
A (100, 120, 220, 230, or 240 VAC monitors)
5.
Record the patient-to-ground leakage for the following conditions:
Table 9-3. Patient-to-Ground Leakage Conditions for ECG
Neutral
Ground
Power (Polarity)
Closed
Closed
Normal
Closed
Open
Normal
Closed
Open
Reversed
Closed
Closed
Reversed
Open
Closed
Normal
Open
Closed
________ Reversed
Содержание Corometrics 126
Страница 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 2: ......
Страница 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Страница 30: ...1 10 120 Series Maternal Fetal Monitor Revision B 2015590 001 Safety Equipment Symbols For your notes ...
Страница 404: ...A 6 120 Series Maternal Fetal Monitor Revision B 2015590 001 Factory Defaults Table of Defaults For your notes ...
Страница 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Страница 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Страница 412: ......
Страница 414: ......
Страница 416: ......
Страница 418: ......
Страница 420: ......
Страница 422: ......
Страница 424: ......
Страница 426: ......
Страница 428: ......
Страница 430: ......
Страница 432: ......
Страница 434: ......
Страница 436: ......
Страница 438: ......
Страница 439: ......