6-12
120 Series Maternal/Fetal Monitor
Revision B
2015590-001
Functional Checkout Procedure: FECG Test
8.
Repeat
step 7
for each of the following rates: 30, 60, 210, and 240 BPM.
9.
Change the simulator’s
QRS Polarity
switch from + to –. Verify that the
monitor does not skip any beats.
10. Set the simulator’s
ECG Rate
switch to the RAMP setting. Verify that the
monitor’s FHR1 value counts between approximately 30 and 240 BPM and that
the recorder prints a ramp between the same values. (Refer to
Figure 6-4
.)
11. Access the Install Options service screen and set ECG Artifact Elimination to
OFF; then exit the service mode.
12. Set the simulator’s
ECG Rate
switch to the
∆
15 position. Verify the following
on the monitor:
The FHR1 value oscillates by 15 BPM.
The FHR1 heartbeat indicator (
) flashes for each input signal.
The ECG “beep” is heard from the rear panel speaker.
The recorder prints an oscillation of 15 BPM between 115 and 130 BPM on
the top grid of the strip chart paper. (Refer to
Figure 6-5
.)
13. Repeat
step 12
for rate values of
∆
22 and
∆
27. The results should be the same
except that the FHR1 value oscillates by either 22 or 27 BPM and the recorder
prints an oscillation of 22 or 27 BPM. The top value is always at approximately
130 BPM. (Refer to
Figure 6-5
.)
14. Access the monitor’s Install Options service mode screen and set the ECG
Artifact Elimination to ON.
Table 6-2. FECG Test Simulator Settings
Section
Switch
Setting
FECG/MECG
Main
RATE
Rate
MANUAL
Mode
FECG
QRS Amplitude
15 µV
QRS Polarity
+
GENERAL
Pattern Memory
OFF
UA
Main
CMR
Mode
TOCO
Содержание Corometrics 126
Страница 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 2: ......
Страница 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Страница 30: ...1 10 120 Series Maternal Fetal Monitor Revision B 2015590 001 Safety Equipment Symbols For your notes ...
Страница 404: ...A 6 120 Series Maternal Fetal Monitor Revision B 2015590 001 Factory Defaults Table of Defaults For your notes ...
Страница 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Страница 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Страница 412: ......
Страница 414: ......
Страница 416: ......
Страница 418: ......
Страница 420: ......
Страница 422: ......
Страница 424: ......
Страница 426: ......
Страница 428: ......
Страница 430: ......
Страница 432: ......
Страница 434: ......
Страница 436: ......
Страница 438: ......
Страница 439: ......