9-10
120 Series Maternal/Fetal Monitor
Revision B
2015590-001
General Maintenance: Maternal Blood Pressure Tests
10. Slowly inflate the manometer, using the inflation hand bulb, to 285 mmHg ± 15
mmHg. Verify that the
OVERPRESSURE
message displays.
WARNING
OVERPRESSURE—If the overpressure message is not indicated,
or if the overpressure reading is not within the specified tolerance,
contact your Service Representative. Do not use this monitor to
perform blood pressure readings until the problem is corrected.
11. Access the Diagnostic Control service mode screen and configure the NBP
mode to PATIENT.
Table 9-1. Calibration Check Pressure Levels
MAnometer Pressure Level
120 Series Monitor Pressure Level
0 mmHg + 1 mmHg
0 mmHg + 1 mmHg
50 mmHg ± 1 mmHg
50 mmHg ± 4 mmHg
100 mmHg ± 1 mmHg
100 mmHg ± 4 mmHg
150 mmHg ± 1 mmHg
150 mmHg ± 4 mmHg
200 mmHg ± 1 mmHg
200 mmHg ± 5 mmHg
Содержание Corometrics 126
Страница 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 2: ......
Страница 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Страница 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Страница 30: ...1 10 120 Series Maternal Fetal Monitor Revision B 2015590 001 Safety Equipment Symbols For your notes ...
Страница 404: ...A 6 120 Series Maternal Fetal Monitor Revision B 2015590 001 Factory Defaults Table of Defaults For your notes ...
Страница 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Страница 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Страница 412: ......
Страница 414: ......
Страница 416: ......
Страница 418: ......
Страница 420: ......
Страница 422: ......
Страница 424: ......
Страница 426: ......
Страница 428: ......
Страница 430: ......
Страница 432: ......
Страница 434: ......
Страница 436: ......
Страница 438: ......
Страница 439: ......